The xDbio XD116 VET 6-Part Hematology Analyzer is a professional, high-precision diagnostic tool engineered exclusively for animal blood cell analysis—developed by xDbio Biotechnology Co., Ltd(a company of BMV). Designed to meet the diverse needs of veterinary hospitals, clinics, animal research centers, livestock diagnosis facilities, and marine life care institutions, this analyzer integrates advanced technologies to deliver comprehensive, accurate blood test results for over 19 animal species. With a focus on precision, efficiency, and user-friendliness, it combines semiconductor laser scatter flow cytometry, nucleic acid fluorescence staining, and sheath flow impedance technology, enabling detailed classification of blood cells and detection of key parameters critical for animal health diagnosis, disease monitoring, and scientific research. Its compact design, intuitive operation, and versatile compatibility make it an indispensable tool for modern veterinary practices seeking reliable hematology analysis solutions.
Core Product Advantages
Advanced Technology for Precise 6-Part Classification
The XD116 VET adopts three core technologies to ensure accurate blood cell analysis:
- Semiconductor Laser Scatter Flow Cytometry: Uses forward scatter (FSC) to measure cell size, side scatter (SSC) to reflect internal cell structure, and side fluorescence (SFL) to detect nucleic acid content—enabling precise differentiation of white blood cell subpopulations.
- Nucleic Acid Fluorescence Staining: A single reagent lyses red blood cells, creates pores in white blood cell membranes, and allows fluorescent dyes to bind to nucleic acids and specific granules (e.g., eosinophils). This ensures high accuracy in cell recognition and classification, with enhanced sensitivity for abnormal cell detection.
- Sheath Flow Impedance Technology: Wraps cells in sheath fluid to ensure they pass through the detection aperture one by one, eliminating pulse distortion and reducing cell clumping interference—greatly improving counting accuracy.
Comprehensive Parameter Coverage & Multi-Species Compatibility
The analyzer detects 35 key hematology parameters, covering four major categories:
- White Blood Cell Parameters (13 items): WBC, NEU (#/%), LYM (#/%), MON (#/%), EOS (#/%), BAS (#/%), IG (#/%), etc.
- Red Blood Cell Parameters (8 items): RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD.
- Reticulocyte Parameters (8 items): RET (#/%), IRF, LFR, MFR, HFR, RET-He, etc.—providing sensitive indicators for iron deficiency anemia and erythropoiesis function.
- Platelet Parameters (7 items): PLT, MPV, PDW, PCT, PLCC, PLCR, PF.It supports 19 animal species (expandable) with species-specific reference ranges, including common pets (dogs, cats, rabbits, ferrets, mini pigs, alpacas), livestock (horses, cattle, pigs, goats, sheep), laboratory animals (gerbils, guinea pigs, hamsters, mice), and marine animals (dolphins, sea lions, seals).
Efficient Operation & User-Centric Design
- High Analysis Speed: Processes 40 samples per hour (CBC+DIFF+RET), meeting the needs of high-volume clinics and research labs.
- Intuitive Interface: Equipped with a 10.4-inch TFT touchscreen for easy operation, with simple menu navigation and clear result display.
- Flexible Sampling & Data Output: Features open sampling design; supports data export via USB/LAN (CSV format, editable in Excel) and A4 report printing via external printers.
- Low Maintenance: Automatically performs cleaning procedures during startup and shutdown, reducing manual maintenance workload.
- Compact & Portable: Weighs 25kg with dimensions of 503×311×414mm, saving laboratory space and facilitating mobility.
Dual Platelet Detection & Abnormal Alert FunctionThe analyzer uses two methods for platelet counting to ensure accuracy:
- Sheath Flow Impedance Method (PLT-I): Offers high accuracy and repeatability for normal samples.
- Optical Method (PLT-O): Automatically switches to this method for abnormal samples (e.g., large platelets, platelet aggregation, red blood cell fragments), eliminating interference and ensuring reliable results.It also features abnormal result marking and alert functions—for example, setting a BAND (band neutrophil) alert zone within the neutrophil counting range to distinguish BAND cells from lymphocytes, reducing misjudgment and prompting re-examination or manual smear counting when necessary.
Reliable Performance & Wide Application Scope
Built with high-quality components and precision manufacturing, the XD116 VET delivers stable performance under a wide operating environment (temperature: 15-30°C, humidity: 30%-85%, atmospheric pressure: 70-105kPa). It is suitable for diverse scenarios:
- Veterinary clinics/hospitals: Routine health checks, disease diagnosis, and treatment monitoring for pets.
- Livestock diagnosis: Health screening and disease control for cattle, pigs, sheep, etc.
- Animal research centers: Laboratory animal hematology analysis for scientific studies.
- Marine life care: Health assessment for dolphins, sea lions, and other marine animals.
| Specification Category | Parameter Details |
| Product Model | xDbio XD116 VET 6-Part Hematology Analyzer |
| Core Technologies | Semiconductor laser scatter flow cytometry, nucleic acid fluorescence staining, sheath flow impedance technology |
| Detection Parameters | – White Blood Cell (13 items): WBC, NEU (#/%), LYM (#/%), MON (#/%), EOS (#/%), BAS (#/%), IG (#/%):- Red Blood Cell (8 items): RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD- Reticulocyte (8 items): RET (#/%), IRF, LFR, MFR, HFR, RET-He- Platelet (7 items): PLT, MPV, PDW, PCT, PLCC, PLCR, PF- Total: 35 parameters (varies by species) |
| Sample Requirements | – Sample Type: EDTA-K2 anticoagulated whole blood- Sample Volume: 25μL |
| Analysis Performance | – Analysis Speed: 40 samples/hour (CBC+DIFF+RET)- Species Compatibility: 19 species (expandable)- Reference Ranges: Species-specific reference ranges provided- Data Storage: 18,000 records (including 4 scatter plots and 2 histograms) |
| Key Functions | – 6-part white blood cell classification- Dual platelet detection (PLT-I/PLT-O)- Reticulocyte analysis- Abnormal result alert and marking- Automatic startup/shutdown cleaning- BAND cell alert zone setting |
| Display & Interface | – Display: 10.4-inch TFT touchscreen- Interfaces: USB, LAN- Data Output: CSV file export, A4 report printing (external printer required) |
| Physical & Environmental | – Dimensions (L×W×H): 503mm×311mm×414mm- Weight: 25kg- Operating Temperature: 15-30°C- Relative Humidity: 30%-85% (non-condensing)- Atmospheric Pressure: 70-105kPa- Power Supply: 100V-240V~, 50/60Hz |
| Consumables & Maintenance | – Reagents: Diluent (DS/DR), lysate (LD/LH), staining solution (FD/FR)- Maintenance Supplies: Probe cleaning solution- Automatic Cleaning: Startup/shutdown automatic cleaning software |