Description
The XD113 is a high-performance fully automatic 3-part hematology analyzer engineered for clinical routine complete blood count (CBC) testing, ideal for hospitals, primary care clinics, clinical laboratories and community health centers. Adopting classical electrical impedance measurement technology and a safe cyanide-free HGB detection method, it combines ultra-low sample volume, high testing efficiency, massive data storage and user-friendly operation with low operational cost. Designed with clinical practicality at its core, it features advanced sweep current technology for accurate platelet counting, a glove-operable touch screen, and seamless LIS integration, delivering reliable, precise 3-part WBC differential results while meeting the diverse testing needs of daily clinical workflow.
Core Product Advantages
Safe & Reliable Detection Technologies with High Accuracy
- Core principles: Classical electrical impedance measurement for WBC, RBC, PLT counting and differentiation; cyanide-free HGB measurement technique eliminates toxic cyanide reagents, ensuring biohazard-free operation for lab staff and environmental protection.
- PLT accuracy guarantee: Advanced sweep current technology specifically optimizes platelet detection, avoiding counting errors caused by small particle interference and ensuring high precision for PLT results—critical for clinical bleeding and thrombotic disease diagnosis.
- Excellent performance: Boasts a wide linear range, ultra-low carryover rate (≤0.5% for WBC/RBC/HGB) and high repeatability (CV ≤2.0% for WBC), delivering consistent and accurate test results for all core hematology parameters.
Ultra-Low Sample Volume & High Testing Efficiency
- Minimal blood consumption: ≤10μL of sample per test, reducing patient sampling pain—perfect for pediatric, geriatric and patients with limited blood volume, and suitable for peripheral blood testing scenarios.
- High throughput: Up to 60 samples per hour, meeting the daily routine testing demands of medium and small clinical facilities, with fast result output to optimize lab workflow efficiency.
- Multiple measurement modes: Supports venous whole blood, pre-diluted peripheral blood and anticoagulated peripheral whole blood modes, flexible for different clinical sampling and sample preparation needs, no external computer required for operation.
User-Friendly Design for Effortless Operation
- Glove-operable 10.4-inch capacitive color touch screen: Ultra-sensitive touch response enables smooth operation even with rubber medical gloves, a key convenience for lab staff in sterile working environments; the high-definition display visualizes 3 histograms and test results clearly.
- One-click operation: One-click testing and one-click troubleshooting eliminate complex operational steps, reducing the learning curve for new users; all daily work information is displayed holistically on a single interface, avoiding tedious multi-interface switching.
- Flexible result output: Equipped with a built-in thermal printer for on-site quick printing, and supports external printer connection; multiple output formats meet different clinical reporting needs, with clear visualization of numeric results and histograms.
Low Operational Cost & High Cost-Effectiveness
- Minimal reagent demand: Only two dedicated reagentsare required for all tests, simplifying reagent management and reducing inventory costs; low single-test reagent consumption further lowers the daily operational cost of the analyzer.
- Durable & low-maintenance: The robust hardware design ensures long-term stable operation with minimal maintenance needs, reducing downtime and repair costs for clinical facilities—ideal for high-frequency daily testing.
Massive Data Storage & Strong Connectivity
- 100,000+ data storage capacity: Stores over 100,000 test results including complete patient information and histograms, supporting long-term historical data query, review and clinical traceability—no need for additional external storage devices.
- Comprehensive connectivity: 4 USB ports, 1 network port, 1 serial port; supports bidirectional LIS system and HL7 protocolfor seamless integration with hospital laboratory information systems, realizing automatic data transmission and management. Optional built-in RFID reader/writer enables intelligent reagent and sample management, further improving lab automation level.
Compact & Adaptable to Diverse Environments
- Compact size (315mm×416mm×411mm) and lightweight (approx. 26kg) save valuable lab space, easy to install and place in small clinics or mobile testing stations.
- Wide environmental adaptability: Universal AC power (100-240V, 50/60Hz), operable at 10-30°C, humidity 20-85%, atmospheric pressure 70-106KPa and altitude ≤3500m—suitable for clinical facilities in different regions and altitudes.
| Specification Category | Parameter Details |
| Detection Principles | WBC/RBC/PLT: Classical Electrical Impedance Measurement;HGB: Cyanide-Free HGB Measurement Technique;PLT: Advanced Sweep Current Technology for Accuracy |
| Reported Parameters | 21 parameters (WBC, Lym#, Mid#, Gran#, Lym%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC) + 3 Histograms |
| Measurement Modes | Venous whole blood, Pre-diluted peripheral blood, Anticoagulated peripheral whole blood |
| Sample Volume | ≤10μL per test |
| Testing Throughput | 60 samples per hour |
| Display | 10.4-inch High-Definition Capacitive Color Touch Screen; Glove-operable, Ultra-sensitive |
| Data Storage | 100,000+ copies (including histograms and complete patient information) |
| Result Output | Built-in thermal printer; External printer support |
| Interfaces | 4×USB ports, 1×Network port, 1×Serial port; Optional built-in RFID reader/writer |
| Communication | Bidirectional LIS system support; HL7 protocol compatible |
| Operating Environment | Voltage: 100-240V a.c. (50/60Hz); Temperature: 10-30°C;Humidity: 20-85%; Atmospheric pressure:70-106KPa; Altitude: ≤3500m |
| Dimension (W×H×D) | 315mm × 416mm × 411mm |
| Weight | Approx. 26kg |
| Core Performance Indices | See Performance Table below |
| Operational Features | One-click testing/troubleshooting; Single-interface holistic information display; No external computer required |
| Reagent Requirement | Only 2 dedicated reagents for all tests; Low single-test reagent consumption |
Performance Parameter Table
| Item | Linear Range | Carryover Rate | Repeatability (CV Value) |
| WBC | 0-200×10⁹/L | ≤0.5% | ≤2.0% |
| RBC | 0-8×10¹²/L | ≤0.5% | ≤1.5% |
| HGB | 0-250g/L | ≤0.6% | ≤1.5% |
| PLT | 0-1000×10⁹/L | ≤1.0% | ≤4.0% |